The experience reached with the analyser BT3000 Plus, one of the most famous and diffused analysers for clinical chemistry in the world, have produced his/her logic evolution:
The point of reference for the future in the modern laboratory of clinical analysis.The Know how thirty-
–
–
–
–
–
still more…
Be the first to review “BT 3000 plus”
| TECHNICAL SPECIFICATIONS | |
| Throughput | Up to 300 Chem. Tests/hr + up to 190 ISE Tests/hr |
| ISE Module on board | Yes: Na, K, Cl, CO2 |
| Number of reagents | 80 Total – (40 for or 80ml + 40 for 10ml or 20ml) |
| Refrigerated Reagent Bay | Yes |
| Reaction volume | From 280 to 700 µl |
| Sample volume | From 1 to 100 µl |
| Reading cuvettes | 34 non-disposable optical glass |
| Sampling Arm | 2 for Serum and Reagents |
| Number of position | 78 Total – 52 Samples – 26 Standards and Control |
| Test Modes | Random Access, Batch, STAT |
| Test Directory | Up to 500 |
| Temperature control | Peltier System OFF/30/32/37°C (±0,2°C) |
| Vacuum Pump | External |
| Photometer | Solid state 10 channel + Reference channel |
| Photometer Accuracy | ± 1% (0-2 O.D.), + 2,5% (2-3 O.D.) |
| Photometer Sensitivity | ± 0,001 Abs |
| Diluter | 2 – Long life ceramic 470 µl |
| Diluter Accuracy | ± 0,1% f.s. |
| Diluter Sensitivity | ± 1 µl |
| Bar Code | 2: Reagent and Serum |
| Internal Computer | CPU Atom 1.8 GHz |
| Peripheral Devices | DVD ROM, HD>40 Gb |
| Monitor | Color LCD 12″ |
| Touch-screen | Yes: Glass Protected resistive kind |
| Keyboard | U.S.A. Standard |
| Mouse | Yes |
| Software | Windows® 7 32 bit |
| Process electronic | 12 Microcontrollers I² CBUS Protocol |
| Interface | RS232, Bidirectional + USB + LAN + VGA |
| Built in QC Program | Yes: 3 Levels known, 3 Levels Unknown |
| Maintenance | Software assisted |
| Power Supply | 100/240 Volt, 50-60Hz, 1000 Watts, PFC |
| Environmental temperature | 18 – 32°C (64.4-89.6°F) |
| Relative humidity | 10 – 85% (not condensing) |
| Dimension | Height – 27″, Width – 40″, Depth – 23″ |
| Weight | 210 lb (95 Kg.) |
| Windows 2000®, Intel Pentium® and Cubitainer® are registered of Microsoft Corporation, Intel and Emergency Resources. IBM is registered trademark of the International Business Machine. | |
| Features may change without notice | |